Artificial organs
-
The objective was to assess the i-cor electrocardiogram-synchronized diagonal pump in terms of hemodynamic energy properties for off-label use in neonatal and pediatric extracorporeal life support (ECLS) circuits. The neonatal circuit consisted of an i-cor pump and console, a Medos Hilite 800 LT oxygenator, an 8Fr arterial cannula, a 10Fr venous cannula, 91 cm of 0.6-cm ID arterial tubing, and 91 cm of 0.6-cm ID venous tubing. The pediatric circuit was identical except it included a 12Fr arterial cannula, a 14Fr venous cannula, and a Medos Hilite 2400 LT oxygenator. ⋯ Under pulsatile mode, preoxygenator SHE increased with increasing postarterial cannula pressure and pulsatile amplitude, but decreased with increasing flow rate. The i-cor system can provide nonpulsatile and pulsatile flow for neonatal and pediatric ECLS. Pulsatile amplitudes of 2000-3000 rpm are recommended for use in neonatal and pediatric patients.