Pacing and clinical electrophysiology : PACE
-
Pacing Clin Electrophysiol · Aug 1989
Randomized Controlled Trial Comparative Study Clinical TrialActive fixation atrial leads: randomized comparison of two lead designs.
Active fixation leads have reduced the incidence of lead dislodgement in patients with permanent pacemakers. However, theoretic concern that the tissue trauma associated with a myocardial screw-helix may increase the chronic pacing threshold of active compared to passive fixation leads has remained. Whether active fixation leads with a stimulating electrode that is independent of the fixation mechanism are associated with a lower chronic pacing threshold than leads utilizing a screw-helix for both fixation and stimulation is unknown. ⋯ At 6 weeks follow-up, there were no differences in the mean threshold voltage (1.85 +/- 0.36 vs 1.93 +/- 0.69 V), impedance (528 +/- 81 vs 530 +/- 118 ohms), or atrial electrogram amplitude (2.63 +/- 0.50 vs 2.42 +/- 0.95 mV) between the two leads. At long-term follow-up (mean 16.2 +/- 2.8 months, range 13.1-20.0 months) there were no significant differences in voltage threshold (1.65 +/- 0.61 vs 1.97 +/- 0.64 V), impedance (565.5 +/- 81.6 vs 617.7 +/- 146.7 ohms), or atrial electrogram amplitude (2.79 +/- 0.75 vs 3.10 +/- 1.53 mV). Thus, these results suggest that active fixation leads in the atrium with an electrode that is independent of the fixation mechanism do not provide chronic stimulation thresholds or electrogram amplitudes that are superior to those obtained with leads utilizing a myocardial screw-helix as both the active electrode and the fixation device.