Comparative StudyAtrial and ventricular stimulation threshold development: a comparative study in patients with a DDD pacemaker and two identical carbon-tip leads.
The atrial and ventricular pacing threshold development during the first postoperative year was studied in a group of patients receiving DDD pacemakers. Identical carbon-tip endocardial leads were implanted in atrium and ventricle. Atrial and ventricular voltage stimulation thresholds were measured at implantation, and noninvasively at 1 and 12 months thereafter. ⋯ The ratio of chronic to acute stimulation thresholds was higher on the ventricular than on the atrial level (0.001 greater than P greater than 0.0005). The chronic atrial threshold showed a logarithmic relation to the threshold at implantation (P = 0.0006); postoperative threshold rise was not a significant determinant of the chronic atrial threshold (P = NS). On the ventricular level, the reverse was seen: The chronic threshold was related to the postoperative threshold rise (P = 0.0015, logarithmic relation), but not to the implantation threshold (P = NS).(ABSTRACT TRUNCATED AT 250 WORDS)