Comparative StudyComparison of impedance minute ventilation and direct measured minute ventilation in a rate adaptive pacemaker.
Respiration rate (RR) and minute ventilation (MV) provide important clinical information on the state of the patient. This study evaluated the accuracy of determining these using a pacemaker impedance sensor. In 20 patients who were previously implanted with a Guidant PULSAR MAX group of pacemakers, the telemetered impedance sensor waveform was recorded simultaneously with direct volume respiration waveforms as measured by a pneumatometer. ⋯ The correlation coefficient for iMV versus MV over the entire 10 minutes of exercise, including the initial 4 minutes of exercise, was 0.99. The transthoracic impedance sensor of an implanted pacemaker can accurately detect respiration parameters. There was a large variation between subjects in the iMV versus MV slope during a bicycle exercise test, whereas for each subject, the slope was stable during submaximal bicycle exercise.