International journal of pharmaceutics
-
Randomized Controlled Trial Comparative Study Clinical Trial
In Check Dial: accuracy for Diskus and Turbuhaler.
The In Check Dial was developed to evaluate whether a patient is able to generate an adequate peak inspiratory flow (PIF) through a certain inhalation device. The inhalation profile recorder (IPR) is a calibrated instrument that measures flows through Diskus and Turbuhaler in our research setting. The aim of this study was to compare the PIFs of patients when inhaling through a Diskus or Turbuhaler connected with the IPR (PIF_diskus and PIF_TH) to the flows through the corresponding orifices of the In Check Dial (Diskus_In Check and TH_In Check). ⋯ Measuring PIF through Diskus and Turbuhaler using the IPR and the In Check Dial, respectively shows a disagreement of 3.9l/min. A disagreement of 3.5l/min was found for the Turbuhaler. The In Check Dial did not identify two of four patients as 'non-optimal' users.