Comparative StudySequential or single pulse defibrillation? Investigations towards energy reduction in experimental animals.
With introduction of the automatic implantable cardioverter-defibrillator for treatment of medically refractory ventricular arrhythmias, many investigations are focussing on possibilities of reducing the energy necessary for defibrillating the heart to obtain a more adequate size and a longer durability of the generator. Several studies favour the sequential pulse delivery, using three electrodes, either endocardial, epicardial or subcutaneous plates, to improve defibrillation performance of low energy shocks. However, the validity of this conclusion remains equivocal since two different electrode configurations were used for single and sequential defibrillation. ⋯ Therefore, we have to discuss a different electrophysiological mechanism for sequential defibrillation than for single shocks with an extended current distribution over the heart using four electrodes. Possibly the action potential of the single cell of the fibrillating heart is of interest in reflecting the efficiency of sequential pulse defibrillation. By influencing the action potential pharmacologically in some animals, the optimal time interval for double shocks could be extended up to 7 msec.(ABSTRACT TRUNCATED AT 250 WORDS)