[HYGEA (Hygiene in gastroenterology--endoscope reprocessing): Study on quality of reprocessing flexible endoscopes in hospitals and in the practice setting].
The quality of reprocessing gastroscopes, colonoscopes and duodenoscopes in daily routine of 25 endoscopy departments in hospitals and 30 doctors with their own practices was evaluated by microbiological testing in the HYGEA interventional study. In 2 test periods, endoscopes ready for use in patients were found contaminated at high rates (period 1: 49 % of 152 endoscopes; period 2: 39 % of 154 endoscopes). Culture of bacterial fecal flora (E. coli, coliform enterobacteriaceae, enterococci) was interpreted indicating failure of cleaning procedure and disinfection of endoscopes. ⋯ Bottle and tube for air/water-channel flushing have to be reprocessed daily by disinfection or sterilization, and in use, the bottle have to be filled exclusively with sterile water. The HYGEA study shows that microbiological testing of endoscopes is useful for detection of insufficient reprocessing and should be performed for quality assurance in doctors' practices, too. The study put recommendations for reprocessing procedures in more concrete terms.