Annals of clinical biochemistry
-
Ann. Clin. Biochem. · Mar 2005
Measurement of urinary free cortisol using liquid chromatography-tandem mass spectrometry: comparison with the urine adapted ACS:180 serum cortisol chemiluminescent immunoassay and development of a new reference range.
The measurement of urinary free cortisol (UFC) is commonly used in the investigation of possible Cushing's syndrome. With the recent availability of liquid chromatography-tandem mass spectrometry (LC-MS/MS) in hospital laboratories, we wanted to develop a specific UFC LC-MS/MS method and compare it with our current immunoassay method and develop a new LC-MS/MS reference range if required. ⋯ We have developed a UFC LC-MS/MS method with a solid-phase extraction clean-up step. The method shows adequate performance and is suitable for routine laboratory use. The mixed sex (n = 110, men = 25, women = 85) reference range was up to 165 nmol/24 h or 146 nmol/24 h, depending on how the data are manipulated.