Annals of clinical biochemistry
-
Ann. Clin. Biochem. · Sep 2018
Valid analytical performance specifications for combined analytical bias and imprecision for the use of common reference intervals.
Background Many clinical decisions are based on comparison of patient results with reference intervals. Therefore, an estimation of the analytical performance specifications for the quality that would be required to allow sharing common reference intervals is needed. The International Federation of Clinical Chemistry (IFCC) recommended a minimum of 120 reference individuals to establish reference intervals. ⋯ The formulae are identical for Gaussian and log-Gaussian distributions. Results Method 2 gives the correct results with a constant percentage of 4.4% for all combinations of bias and imprecision. Conclusion The Microsoft Excel formula NORMINV is useful for the estimation of analytical performance specifications for both Gaussian and log-Gaussian distributions of reference intervals.