Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models.
To compare (1) pullout properties between 3.5 mm cortical and locking screws, and (2) mechanical properties and gap displacements between the 3.5 mm broad limited-contact dynamic compression plate (LC-DCP), broad dynamic compression plate (DCP), and narrow locking compression plate (LCP), during axial loading of plate-stabilized diaphyseal fragments with an interfragmentary gap. ⋯ The LCP system is more likely than LC-DCP and DCP systems, with neutrally positioned screws, to maintain a planned interfragmentary gap, although gap strains range from 0% to 15% across the 2 mm gap during a trot load.