Case Reports[Interference between a dual-chamber pacemaker and argon electrocautery device during hepatectomy].
We report an unusual electromagnetic interference induced by an argon electrocautery device during a left hepatectomy on a dual chamber pacemaker, implanted for sinus node dysfunction in 87-year-old patient. Argon electrocautery induced inhibition of atrial stimulation and occurrence of irregular ventricular triggered activity. Normal pacemaker function resumed after electrocautery interruption. This case illustrates the need to focus on cardiac rhythm when a new electrical device is used in a pacemaker patient.
-
To review fresh frozen plasma (FFP) prescriptions and compare their validity to the legal french guidelines (law of the 12/03/91). ⋯ A significant proportion of FFP transfusions is inappropriate. This study, which is the first step of a quality assurance program, will be followed by local recommendations for clinical practice. The current standards of prescribing FFP are more restrictive than those defined in the legal french guidelines.