Journal of hypertension
-
Journal of hypertension · Aug 1989
Review Comparative StudyOptimal size of cuff bladder for indirect measurement of arterial pressure in adults.
This study tested the hypothesis that a sphygmomanometer cuff bladder long enough to encircle the arm in most adults ('obese cuff') would provide a more accurate and precise estimate of intra-arterial pressure than the usual 'standard' cuff bladder. In 53 patients undergoing diagnostic coronary angiography (35 males, 18 females, aged 36-79 years), indirect blood pressure, measured in the left arm with a random-zero sphygmomanometer, was compared with simultaneously measured femoral intra-arterial pressure. Duplicate indirect measurements were made with each of two cuffs containing bladders measuring 39 x 15 cm ('obese') and 23 x 12 cm ('standard'). ⋯ Both cuffs underestimated systolic blood pressure, the obese cuff by 15.5 +/- 11.7 mmHg and the standard cuff by 7.6 +/- 12.1 mmHg. These systolic blood pressure underestimates were greater at higher blood pressures (P less than 0.01) and with smaller arms (P less than 0.05). Age was not related to measurement error with either cuff.(ABSTRACT TRUNCATED AT 250 WORDS)