Comparative StudyNon-invasive assessment of local arterial pulse pressure: comparison of applanation tonometry and echo-tracking.
Pulse pressure is not constant throughout the arterial tree. Use of pulse pressure at one arterial site as surrogate for pulse pressure at another arterial site may be erroneous. The present study compares three non-invasive techniques to measure local pulse pressure: (i) internally calibrated readings from applanation tonometry, (ii) alternative calibration of pressure waves obtained with applanation tonometry and (iii) alternative calibration of arterial distension waves obtained with echo-tracking. Alternative calibration assumes mean and diastolic blood pressure constant throughout the large artery tree. ⋯ Pulse pressure obtained from alternatively calibrated tonometer-derived pressure waves and echo-tracking-derived distension waves demonstrates good accuracy. Accuracy of pulse pressure from internally calibrated applanation tonometer readings at the carotid artery is poor.