Journal of pharmaceutical and biomedical analysis
-
J Pharm Biomed Anal · Jul 2007
A rapid and simple HPLC method for the analysis of propofol in biological fluids.
A selective and sensitive high-performance liquid chromatographic method for the analysis of propofol in biological samples was developed. Propofol and thymol (internal standard) were analysed on a Purospher RP-18 endcapped (75 mmx4 mm, 3 microm) stationary phase using acetonitrile and water (65:35, v/v) as eluents at a flow rate of 0.6 mL/min. The excitation and emission wavelengths were 276 and 310 nm, respectively. ⋯ The intra-day and inter-day precisions were lower than 5.5% for three concentrations assessed (0.05, 1.0 and 10.0 mg/L). Considering the column size and the flow rate, the separation was achieved with an analysis time less than 6 min with a reduced consumption of solvent. This rapid HPLC method using a simple treatment procedure is sensitive enough for monitoring propofol in human biological samples.