IEEE transactions on medical imaging
-
IEEE Trans Med Imaging · Dec 2004
Comparative Study Clinical TrialVascular segmentation of phase contrast magnetic resonance angiograms based on statistical mixture modeling and local phase coherence.
In this paper, we present an approach to segmenting the brain vasculature in phase contrast magnetic resonance angiography (PC-MRA). According to our prior work, we can describe the overall probability density function of a PC-MRA speed image as either a Maxwell-uniform (MU) or Maxwell-Gaussian-uniform (MGU) mixture model. An automatic mechanism based on Kullback-Leibler divergence is proposed for selecting between the MGU and MU models given a speed image volume. ⋯ It is shown that segmentation based on both measures gives a more accurate segmentation than using either speed or flow coherence information alone. The proposed method is tested on synthetic, flow phantom and clinical datasets. The results show that the method can segment normal vessels and vascular regions with relatively low flow rate and low signal-to-noise ratio, e.g., aneurysms and veins.