Revue des maladies respiratoires
-
Compared with invasive ventilation, non-invasive ventilation (NIV) has two unique characteristics: its non-hermetic nature and the fact that the ventilator-lung assembly cannot be considered as a single-compartment model because of the presence of variable resistance represented by the upper airways. When NIV is initiated, the ventilator settings are determined empirically based on clinical evaluation and blood gas variations. However, NIV is predominantly applied during sleep. ⋯ Ventilatory modality, mode of triggering, pressurization slope, use or not of positive end expiratory pressure and type of exhalation as well as ventilator performances may all have physiological consequences. Leaks and upper airway resistance variations may, in turn, modify these patterns. This article discusses the equipment available for NIV, analyses the effect of different ventilator modes and settings and of exhalation and connecting circuits on ventilatory traces and gives the background necessary to understand their impact on nocturnal monitoring of NIV.