The indirect measurement of pleural pressure in neonates is obtained from measurements of esophageal pressure (Pes) with either a liquid-filled catheter or an esophageal balloon-catheter system. The purpose of this investigation was to assess the validity of the water-filled esophageal catheter by comparing the simultaneous changes in Pes and airway opening pressure (Pao) during occluded respiratory efforts. ⋯ During occlusions, changes in Pes and Pao were almost identical in magnitude and timing in each body position studied (right lateral, prone, and supine). We conclude that the water-filled 8-FG esophageal catheter gives an accurate measurement of pleural pressure changes in healthy neonates.