Journal of applied physiology
-
Current practice of monitoring lung ventilation in neonatal intensive care units, utilizing endotracheal tube pressure and flow, end-tidal CO2, arterial O2 saturation from pulse oximetry, and hemodynamic indexes, fails to account for asymmetric pathologies and to allow for early detection of deteriorating ventilation. This study investigated the utility of bilateral measurements of chest wall dynamics and sounds, in providing early detection of changes in the mechanics and distribution of lung ventilation. Nine healthy New Zealand rabbits were ventilated at a constant pressure, while miniature accelerometers were attached to each side of the chest. ⋯ Side identification of the pneumothorax was achieved at 50% tPTX, within a 95% confidence interval. Diagnosis was, on average, 34.1 ± 18.8 min before tPTX. In conclusion, bilateral monitoring of the chest dynamics and acoustics provide novel information that is sensitive to asymmetric changes in ventilation, enabling early detection and localization of pneumothorax.