Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine
-
A theoretical algorithm for shaping the signal response during the approach to steady state in three-dimensional magnetization-prepared rapid gradient-echo (3D MP-RAGE) pulse sequences has been developed and implemented. This algorithm derives the flip angle series required to produce specifically chosen time evolutions of the signal intensities during the data acquisition segment of 3D MP-RAGE sequences. Theoretical predictions for the cases of unshaped, uniform, and mono-exponential decay signal responses were quantitatively validated with a doped-water phantom on a 1.5-T whole-body imager and in all cases there was excellent agreement between the theoretical and experimental values. ⋯ Images obtained from a healthy volunteer using this variable flip angle sequence were compared with 3D RF-spoiled steady-state gradient-echo images obtained in the same total imaging time. The 3D MP-RAGE images demonstrated a 41% increase in the WGSD-to-noise ratio. These initial very promising results indicate that with further refinement to eliminate some intensity artifacts, the variable flip angle 3D MP-RAGE technique may, with respect to certain image properties, provide considerable improvements over currently available 3D gradient-echo imaging techniques.