A new magnetization preparation and image acquisition scheme was developed to obtain high-resolution brain images with optimal tissue contrast. The pulse sequence was derived from an optimization process using simulated annealing, without prior assumptions with regard to the number of radiofrequency (RF) pulses and flip angles. The resulting scheme combined two inversion pulses with the acquisition of three images with varying contrast. ⋯ The use of three-dimensional (3D) sensitivity-encoded (SENSE) echo-planar imaging (EPI) for image acquisition made it possible to achieve a 1.15(3) mm3 isotropic resolution within a scan time of 10 min 21 s. The cortical GM signal-to-noise ratio (SNR) in the calculated GM-only image varied between 30 and 100. The novel technique was evaluated in combination with blood oxygen level-dependent (BOLD) functional magnetic resonance imaging (fMRI) on human subjects, and provided for excellent coregistration of anatomical and functional data.