Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine
-
Although arterial spin labeling (ASL) MRI has been successfully applied to measure gray matter (GM) perfusion in vivo, accurate detection of white matter (WM) perfusion has proven difficult. Reported literature values are not consistent with each other or with perfusion measured with other modalities. In this work, the cause of these inconsistencies is investigated. ⋯ Furthermore, simulations show that even at a spatial resolution of 7 microl voxel size, contamination by the GM signals can exceed 40% of the actual WM signal. From 10-min long flow-sensitive alternating inversion recovery ASL (FAIR-ASL) measurements at 3T in normal subjects (N=7), using highly sensitive detectors, it is shown that single-voxel (7 mul) deep WM perfusion values have an signal-to-noise ratio (SNR) less than 1. The poor sensitivity and heterogeneous transit time limit the applicability of ASL for measurement of perfusion in WM.