Pediatric surgery international
-
Pediatr. Surg. Int. · Jan 2016
Minimal invasive lung support via umbilical vein with a double-lumen cannula in a neonatal lamb model: a proof of principle.
Acute respiratory distress syndrome, with the need for invasive mechanical ventilation (MV) remains a major cause of neonatal mortality and morbidity. Although venovenous extracorporeal lung support (VV-ECLS) has become a standard of care procedure in neonatal patients with acute pulmonary failure there are no reports regarding the use of a double-lumen cannula for extracorporeal minimal invasive lung support via the umbilical vein. ⋯ Even without mechanical ventilation we were able to sufficiently remove pCO2 with our new minimal invasive VV-ECLS using a double-lumen catheter via the umbilical vein, supporting the idea of a lung protective strategy in neonatal acute respiratory failure. pO2 was measured 22, 19 and 23 mmHg, respectively, at the end of VV-ECLS, at least partially caused by recirculation phenomenon, which could possibly be improved by different cannula design. Inotropic support was necessary during VV-ECLS to achieve targeted MAD > 35 mmHg. While technically feasible, this new approach might allow further research in the field of extracorporeal lung support and therefore will follow the concept of a lung protective strategy in acute neonatal respiratory failure.