Biomedical chromatography : BMC
-
Biomed. Chromatogr. · Nov 2006
Liquid chromatographic-electrospray tandem mass spectrometric determination of clarithromycin in human plasma.
A rapid, sensitive and specific LC-MS-MS method has been developed for the determination of clarithromycin (CLA) in human plasma using roxithromycin (ROX) as the internal standard. Samples were prepared via liquid-liquid extraction with methyl tert-butyl ether (MTBE) and chromatographed on a Supelco RP(18) (4.6 x 50 mm, 3 microm particle size) column with a mobile phase consisting of acetonitrile:methanol:60 mM (pH 3.5) ammonium acetate buffer (32.5:32.5:35) at a constant flow rate of 0.8 mL/min. The run time was 3 min with retention times of approximately 1.65 and 1.70 min for CLA and ROX, respectively. ⋯ No matrix ionization suppression was observed when extracted blank matrix or reconstitution solvent was injected onto the system with post-column infusion of clarithromycin and roxithromycin. No carryover was observed when an extracted blank plasma sample was injected immediately after a 5000 ng/mL ULOQ (the upper limit of quantification) standard. The mean recovery was 81.5 and 78.3%, respectively, for clarithromycin and internal standard.