Highly sensitive and rapid ultra-performance liquid chromatography-tandem mass spectrometry method for the determination of nifedipine in human plasma and its application to a bioequivalence study.
An ultra performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method has been developed for the determination of nifedipine in human plasma using nifedipine-d6 as the internal standard (IS). The plasma samples were prepared by solid-phase extraction on Phenomenex Strata-X cartridges employing 200 μL human plasma. Chromatography was carried out on Waters Acquity UPLC BEH C₁₈ (50 × 2.1 mm, 1.7 µm particle size) analytical column under isocratic conditions using a mobile phase consisting of 4.0 mm ammonium acetate-acetonitrile (15:85, v/v). ⋯ Matrix effect was assessed by post-column analyte infusion and the mean extraction recovery was 95.6% across four quality control levels. The method is rugged and rapid with a total run time of 1.2 min and was applied to a bioequivalence study of 20 mg nifedipine tablet formulation in 30 healthy Indian subjects under fasting condition. Assay reproducibility was confirmed by reanalysis of 116 incurred samples.