Perfusion
-
Randomized Controlled Trial Comparative Study Clinical Trial
A clinical evaluation of platelet function, haemolysis and oxygen transfer during cardiopulmonary bypass comparing the Quantum HF-6700 to the HF-5700 hollow fibre membrane oxygenator.
The continued improvement of oxygenators is an important aspect of patient safety during cardiopulmonary bypass (CPB). The purpose of this study was to compare the Bard William Harvey HF-5700 oxygenator to the upgraded Bard Quantum HF-6700, which has recently been introduced into clinical practice. No clinical evaluation of this device has been published to date. ⋯ Samples were obtained at the start of CPB, at 30 min, 60 min, at the end of CPB and at 1 h following termination of CPB. No significant differences between the two groups were found in oxygen transfer, haemolysis (plasma haptoglobin levels) or platelet function (a novel platelet activating factor (PAF)-induced platelet activation test) at any of the time points during CPB. It was concluded that the Quantum HF-6700 matches the HF-5700 for the parameters studied, whilst having the advantage of requiring a smaller priming volume.