Perfusion
-
Randomized Controlled Trial Comparative Study Clinical Trial
The effect of oxygenator membranes on blood: a comparison of two oxygenators in open-heart surgery.
Open-heart surgery (OHS) requires cardiopulmonary bypass (CPB) in most patients. Membrane oxygenators are a critical component of the CPB system. Despite advancements in CPB technology, injury to blood components during CPB still occurs and may result in complications after surgery. ⋯ The Affinity NT oxygenator had a significantly lower difference in pressure across the membrane (p < 0.001) compared with the Optima XP. In conclusion, the two oxygenators performed similarly with respect to their impact on blood components, O2 transfer, and blood loss postoperatively during OHS with CPB. The Affinity NT had the smaller transmembrane pressure drop of the two.
-
Randomized Controlled Trial Comparative Study Clinical Trial
Carbon dioxide field flooding versus mechanical de-airing during open-heart surgery: a prospective randomized controlled trial.
Removal of intracardiac air during valvular surgery should be accomplished in the most effective manner. We conducted a prospective randomized controlled trial to compare mechanical de-airing and carbon dioxide (CO2) field flooding in 18 patients undergoing elective valvular surgery. ⋯ Both assessors graded the bubble count higher in the mechanical deairing group compared with the CO2 flooding group, and there was good agreement between assessors. CO2 field flooding is more effective than mechanical de-airing in removing intracardiac bubbles following valvular surgery.