Perfusion
-
Clinical Trial
Red blood cell transfusion volume and mortality among patients receiving extracorporeal membrane oxygenation.
Red blood cell (RBC) transfusion is used in the critically ill with low hemoglobin concentrations to optimize oxygen utilization and delivery imbalance. Data suggest that RBC transfusion is also independently associated with significant morbidity. We seek to characterize RBC transfusion volumes among patients receiving extracorporeal membrane oxygenation (ECMO) support and test the hypothesis that red blood cell transfusion volume is an independent risk factor for mortality. ⋯ Greater red blood cell transfusion volumes among patients supported with ECMO for non-cardiac indications are independently associated with an increase in odds of mortality. A prospective investigation of restrictive RBC transfusion practices while receiving ECMO may be warranted in this population.
-
Comparative Study
A pilot study comparing two polymethylpentene extracorporeal membrane oxygenators.
We compared two polymethylpentene oxygenators being used in our unit: the Maquet Quadrox-iD paediatric and the Medos Hilite 800LT. ⋯ According to our study, it seems to us that Medos Hilite 800LT membrane oxygenators demonstrate greater oxygen transfer and CO2 removal capacity than Maquet Quadrox-iD paediatric membrane oxygenators, at a similar cost. These results lead us to conclude that it is reasonable to continue using Medos Hilite 800LT membrane oxygenators. A broader comparison study would be necessary in order to support these initial results.
-
Comparative Study
Evaluation of the new generation dual-lumen catheter for neonatal ECMO.
The purpose of this study was to compare the newly designed dual-lumen venovenous catheter (VR13, OriGen Biomedical, Austin, TX) with the current dual-lumen catheter (VV12, OriGen Biomedical). ⋯ These findings indicate that the VR13 catheter resulted in levels of recirculation equal to the VV12. Based on resistance measurements, we do not recommend the use of this new catheter beyond 400 ml/min until minor design changes are made.
-
Comparative Study
In vitro comparison of two different methods of cell washing.
The storage of red blood cells (RBC) results in increased concentrations of plasma free hemoglobin, potassium, glucose, and lactate, among other undesirable substances. These concentrations continue to increase as RBC products age and can cause deleterious effects to the patient. In the setting of cardiac surgery, the autotransfusion devices are routinely used to wash blood that is shed from the surgical site. ⋯ When compared to the original untreated RBCs, the glucose, lactate, and potassium levels were decreased when washed in an autotranfusion device. Additionally, the free hemoglobin and LDH levels were significantly lower with the Medtronic Autolog cell saver than in the COBE 2991 Cell Processor. Washing the RBC donor units in an autotransfusion device prior to transfusion can effectively attenuate the increases seen in glucose, potassium, free hemoglobin, and LDH associated with RBC storage lesion.