[Transcutaneous measurement of partial pressure of carbon dioxide and oxygen saturation: validation of the SenTec monitor].
To validate a monitor for transcutaneous measurement of oxygen saturation (SpO2) and partial pressure of carbon dioxide (TcPCO2). ⋯ The stabilization time recommended for the SenTec monitor before taking a reading is 20 minutes. The overestimates and underestimates by the monitor are not clinically relevant. Finally, the values for SpO2 and TcPCO2 measured by the validated monitor are reliable.
-
Arch. Bronconeumol. · May 2006
[Usefulness of transcutaneous carbon dioxide pressure monitoring to measure blood gases in adults hospitalized for respiratory disease].
To evaluate the usefulness of transcutaneous carbon dioxide pressure (TcPCO2) monitoring in patients hospitalized for respiratory disease. ⋯ Transcutaneous determination of carbon dioxide pressure and oxygen saturation is useful for patients hospitalized for respiratory disease in view of its good correlation and agreement, although SpO2 does tend to overestimate SaO2.