The Journal of arthroplasty
-
Filtered exhaust helmets (space suits) have been reported as efficacious in decreasing wound infections following joint arthroplasty procedures; however, strict experimental control of the many variables related to the incidence of clinical infection has been lacking in these studies, making support of such conclusions difficult. Any potential benefit of filtered exhaust helmets in the control of infection rates can be logically assigned to the reduction of airborne bacterial counts within the operating room. A study was done using the Steri-Shield helmet (Bio-Medical Devices, Irvine, CA) to define its efficacy in limiting airborne bacterial contaminates during total joint arthroplasty. ⋯ Air samples were obtained with an Anderson two-stage viable particle sampler (Anderson Samplers, Atlanta, GA) placed immediately adjacent to the wound. Quantitative microbial cultures were obtained. The filtered exhaust helmet evaluated in this study provided no increased protection against bacterial contamination in the area of the surgical field than conventional paper hoods and masks.