British journal of neurosurgery
-
Comparative Study
A clinical evaluation of the Codman MicroSensor for intracranial pressure monitoring.
The strain-gauge Codman MicroSensor intracranial pressure (ICP) transducer has shown consistently good laboratory performance. To assess the practical performance of the system in patients following acute brain injury, 10 patients were fitted with a MicroSensor and a second ICP monitor. In five cases this was a fibre-optic transducer and in five cases an intraventricular fluid-filled device. ⋯ Comparison traces of ICP in individual patients show high agreement in timing and size of changes. The unexplained constant offset leads to uncertainty about the true ICP. Treatment decisions are often based upon absolute levels of ICP and patient care may therefore differ depending upon the monitor used.