Annals of biomedical engineering
-
Comparative Study
Nonlinear and frequency-dependent mechanical behavior of the mouse respiratory system.
The assessment of the mechanical properties of the respiratory system is typically done by oscillating flow into the lungs via the trachea, measuring the resulting pressure generated at the trachea, and relating the two signals to each other in terms of some suitable mathematical model. If the perturbing flow signal is broadband and not too large in amplitude, linear behavior is usually assumed and the input impedance calculated. Alternatively, some researchers have used flow signals that are narrow band but large in amplitude, and invoked nonlinear lumped-parameter models to account for the relationship between flow and pressure. ⋯ We then extended this model to include nonlinear resistive and elastic terms. We found that the nonlinear elastic term fit the data better than the linear model or the nonlinear resistance model when amplitudes were large. This model may be useful for detecting overinflation of the lung during mechanical ventilation.