We developed a three-dimensional (3D) registration method to align medical scanner data with histological sections. After acquiring 3D medical scanner images, we sliced and photographed the tissue using, a custom apparatus, to obtain a volume of tissue section images. Histological samples from the sections were digitized using a video microscopy system. ⋯ We used the distance between the model surfaces to indicate the 3D registration error. For four experiments, we measured a registration accuracy of 0.96+/- 0.13 mm (mean+/-SD) from the absolute distance between the M2 and H1 model surfaces, which compares favorably to the 0.70 mm in-plane MR voxel dimension. This suggests that our registration method provides sufficient spatial correspondence to correlate 3D medical scanner and histology data.