Annals of biomedical engineering
-
Minimum "hands-off" intervals during cardiopulmonary resuscitation (CPR) are required to improve the success rate of defibrillation. In support of such life-saving practice, a shock advisory system (SAS) for automatic analysis of the electrocardiogram (ECG) contaminated by chest compression (CC) artefacts is presented. Ease of use for the automated external defibrillators (AEDs) is aimed and therefore only processing of ECG from usual defibrillation pads is required. ⋯ The achieved sensitivity of 90.1% meets the AHA performance goal for noise-free VF (>90%). The specificity of 88.5% for NR and 83.3% for ASYS are comparable or even better than accuracy reported in literature. It is important to note that, the aim of this SAS is not to recommend shock delivery but to advice the rescuers to "Continue CPR" or to "Stop CPR and Prepare for Shock" thus minimizing "hands-off" intervals.