Annals of biomedical engineering
-
Aerosol delivery efficiency during noninvasive ventilation (NIV) is known to be low (~10%) and is associated with poor outcomes of aerosol therapy. The objective of this study was to demonstrate the benefit of redesigning ventilation circuit components using a streamlining approach to improve aerosol delivery during nasal high flow therapy in adults with a conventional-sized aerosol from a mesh nebulizer. The ventilation circuit consisted of a humidifier, mesh nebulizer, mixing T-connector (with 90° angle), 10 mm tubing, and nasal cannula interface. ⋯ With the streamlined designs, the highest emitted dose achieved was >40% for a conventional-sized aerosol at 30 LPM. Streamlined geometries offer an effective method to significantly improve the delivery of aerosols through components of NIV systems. This increase in delivery efficiency is important for new inhaled medications with narrow therapeutic windows, increased costs, or long delivery times.