Annals of biomedical engineering
-
This study sought to compare a new head impact-monitoring device, which is not limited to specific helmet styles, against reference accelerometer measurements. Laboratory controlled impacts were delivered using a linear pneumatic impactor to a Hybrid III headform (HIII) fitted with a football helmet and the impact monitoring device (gForce Tracker-GFT) affixed to the inside of the helmet. Linear regression analyses and absolute mean percent error (MAPE) were used to compare the head impact kinematics measured by the GFT to a reference accelerometer located at the HIII's center of mass. ⋯ Results from the impact testing were used to develop a correction algorithm. The coefficients of determination for all impact parameters improved using the correction algorithm for the GFT (r(2) > 0.97), and the MAPE were less than 14%. The GFT appears to be a suitable impact-monitoring device that is not limited to specific styles of football helmets, however, correction algorithms will need to be developed for each helmet style.