Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography
-
J Am Soc Echocardiogr · Jan 2012
Comparative StudyAutomated quantification of mitral inflow and aortic outflow stroke volumes by three-dimensional real-time volume color-flow Doppler transthoracic echocardiography: comparison with pulsed-wave Doppler and cardiac magnetic resonance imaging.
The aim of this study was to compare the feasibility, accuracy, and reproducibility of automated quantification of mitral inflow and aortic stroke volumes (SVs) using real-time three-dimensional volume color-flow Doppler transthoracic echocardiography (RT-VCFD), with cardiac magnetic resonance (CMR) imaging as the reference method. ⋯ RT-VCFD imaging with an automated quantification algorithm is feasible, accurate, and reproducible for the measurement of mitral inflow and aortic SVs and is superior to manual 2D TTE-based measurements. The rapid and automated measurements make this technique practical in the clinical setting to measure and report SVs routinely where the acoustic window will allow it, which was 86% in our study.