A semi-automated solid-phase extraction liquid chromatography/tandem mass spectrometry method for the analysis of tetrahydrocannabinol and metabolites in whole blood.
Marijuana is one of the most commonly abused illicit substances in the USA, making cannabinoids important to detect in clinical and forensic toxicology laboratories. Historically, cannabinoids in biological fluids have been derivatized and analyzed by gas chromatography/mass spectrometry (GC/MS). There has been a gradual shift in many laboratories towards liquid chromatography/mass spectrometry (LC/MS) for this analysis due to its improved sensitivity and reduced sample preparation compared with GC/MS procedures. ⋯ The limits of detection (LODs) ranged from 0.5 to 3 ng/mL and the limits of quantitation (LOQs) ranged from 2 to 8 ng/mL. The bias and imprecision were determined using a simple analysis of variance (ANOVA: single factor). The results demonstrate bias as <7%, and imprecision as <9%, for all components at each quantity control level.