Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
-
Proc Inst Mech Eng H · Aug 2007
Comparative StudyMeasurement of laxity in the anterior cruciate ligament-deficient knee: a comparison of three different methods in vitro.
The aim of this study was to compare in-vitro measurements of anteroposterior laxity in the anterior cruciate ligament (ACL)-deficient knee using three different methods: an Instron materials-testing machine, then a KT-2000 arthrometer, and finally by Roentgen stereophotogrammetric analysis (RSA). Eight ACL-deficient human cadaver knees were used. Total displacement was measured between 90 N anterior and 90 N posterior tibiofemoral drawer forces at both 20 degrees and 90 degrees knee flexion. ⋯ However, the mean Instron measurements of laxity were significantly (3-4 mm) higher than both RSA and KT-2000 measurements. The clinical methods of RSA and the KT-2000 measurements agreed well but appeared to underestimate tibiofemoral anteroposterior laxity compared with the materials-testing machine. These findings may be helpful in the future comparison of different studies.