Ostomy/wound management
-
Ostomy/wound management · Nov 2012
An in vitro assessment of MRI issues at 3-Tesla for antimicrobial, silver-containing wound dressings.
Although no reports of adverse events have been published to date, the presence of metallic dressing ingredients may present an magnetic resonance imaging (MRI) safety concern for patients using silver-containing wound dressings. The purpose of this in vitro study was to test magnetic field interactions (ie, translational attraction and torque), heating, artifacts, and conductivity (ie, electrical resistance) when using MRI at 3-Tesla for two (nonborder and border) silver-containing wound dressings. The results indicated the dressings displayed no magnetic field interactions (deflection angle 0˚; no torque), and in each case, MRI-related heating effects were at the same levels as the background temperature increases (ie, <1.8˚C). ⋯ To date, only a hydrofiber silver-containing dressing has been tested for MRI safety. Because of potential variances in material characteristics, MRI test results are specific to the dressings tested and cannot be applied to other products. Future studies to define the level of silver concentration in dressings that may pose a hazard for performing an MRI are warranted.