Journal of cardiovascular electrophysiology
-
J. Cardiovasc. Electrophysiol. · Nov 2014
ICH E14-compatible holter bin method and its equivalence to individual heart rate correction in the assessment of drug-induced QT changes.
The Holter bin method evaluates QT interval changes in the presence of heart rate changes without correcting the QT interval. However, the method does not allow time-matched comparisons, thus contradicting available guidance and good practice. We report a modification of the methods that allows time-matched comparisons without any heart rate correction. ⋯ The Holter bin methodology can be modified to make it compatible with the available guidance and with good practice of clinical investigations. The results obtained with the modified Holter bin method are practically the same as with individualized heart rate corrected QT intervals. The close correspondence between the 2 methods demonstrates that the present possibilities of comparing QT interval duration in the presence of experiment-induced heart rate differences are not influenced by methodological artifacts.