Comparative StudyComparison of QT Interval Readings in Normal Sinus Rhythm Between a Smartphone Heart Monitor and a 12-Lead ECG for Healthy Volunteers and Inpatients Receiving Sotalol or Dofetilide.
A variety of medications ranging from antiarrhythmics to psychotropics, as well as conditions such as bradycardia, can prolong the QT interval, increasing the risk for life-threatening arrhythmias. Monitoring the corrected QT interval (QTc) is therefore critical for patient safety. The recent development of smart phone heart monitors (SHM) may allow for easier QTc monitoring. We sought to evaluate the accuracy of an SHM for assessing the QTc, as compared to the standard 12-lead ECG. ⋯ The SHM is accurate in measuring QTc interval in sinus rhythm when compared to 12-lead ECG in healthy volunteers. For patients receiving QT prolonging antiarrhythmics, SHM is capable of detecting QTc prolongation, and lead-I of the SHM is most accurate in measuring the QTc if < 500 milliseconds.