A simplified echocardiographic technique for detecting continuous-flow left ventricular assist device malfunction due to pump thrombosis.
Malfunction of a continuous-flow left ventricular assist device (CF-LVAD) due to device thrombosis is a potentially life-threatening event that currently presents a diagnostic challenge. We aimed to propose a practical echocardiographic assessment to diagnose LVAD malfunction secondary to pump thrombosis. ⋯ The 3 echocardiographic variables of measured changes in LV end-diastolic diameter, aortic valve opening time, and deceleration time of mitral inflow between the lowest (8,000 rpm) and highest pump speed settings (11,000 rpm) during echo-guided pump speed-change testing appear highly accurate in diagnosing device malfunction in the setting of pump thrombosis among patients supported with CF-LVAD. Further investigation is warranted to create and validate a prediction score.