Journal of neuroimaging : official journal of the American Society of Neuroimaging
-
Multicenter Study
Preliminary MRI quality assessment and device acceptance guidelines for a multicenter bioclinical study: the GO Glioblastoma Project.
It is a major challenge to guarantee homogeneous acquisition during a prospective multicenter magnetic resonance imaging (MRI) study that makes use of different devices. The goal of the multicenter Grand Ouest Glioblastoma Project (GOGP) was to correlate MRI quantitative parameters with biological markers extracted from image-guided biopsies. Therefore, it was essential to ensure spatial coherence of the parameters as well as the signal intensity and homogeneity. ⋯ A fourth test-object was used to assess T1 accuracy. Although geometry-related parameters, signal-to-noise ratio, uniformity, and T1 measurements varied slightly depending on the different devices, they nevertheless remained within the recommendations and expectations of the multicenter project. This kind of quality control procedure should be undertaken as a prerequisite to any multicenter clinical project involving quantitative MRI and comparison of data acquisitions with quantitative biological image-guided biopsies.