Journal of magnetic resonance imaging : JMRI
-
J Magn Reson Imaging · Jun 2001
Comparative StudyT2-weighted MRI of the female pelvis: comparison of breath-hold fast-recovery fast spin-echo and nonbreath-hold fast spin-echo sequences.
In 49 patients who had pelvic abnormalities, breath-hold T2-weighted fast-recovery (FR)-fast spin-echo (FSE) (imaging time = 24 sec) and nonbreath-hold FSE MR images (2 min 8 sec) were compared qualitatively (on a four-point scale) and quantitatively (using signal-to-noise ratios (SNRs) and contrast ratios (/SIs of the lesions-SIs of the myometrium/SIs of the myometrium)). Motion artifacts were reduced on breath-hold FR-FSE (3.8:3.2 = breath-hold FSE:nonbreath-hold FSE, P < 0.01) and image quality was comparable (3.8:3.7, NS). In all patients, pathology (leiomyoma [N = 26], adenomyosis [N = 10], endometrial carcinoma [N = 8], and ovarian cystic lesions [N = 21]) was recognized with comparable lesion conspicuity (3.8:3.7, NS) and better delineation of the structures (3.9:3.6, P < 0.05) on the FR-FSE images. ⋯ Magn. Reson. Imaging 2001;13:930-937.