Annals of hematology
-
Annals of hematology · Nov 2008
Feasibility, cost-effectiveness and patients' acceptance of point-of-care INR testing in a hospital-based anticoagulation clinic.
Point-of-care (POC) coagulometers are increasingly used by patients for self-monitoring of oral vitamin K antagonists therapy. We studied the feasibility of introducing POC international normalised ratio (INR) testing in place of standard laboratory assays in a hospital-based anticoagulation clinic with 250 active patients. The CoaguChek XS system was first validated in 253 INR samples and found to have a correlation of r = 0.945 with standard assays. ⋯ Among 232 respondents surveyed, 87.5% (95% CI 82.5-91.5%) preferred POC INR monitoring. Rates of thrombosis and major bleeding complications were 1.2% (95% CI 0.2-3.5%) and 0.4% (95% CI 0.01-2.2%), respectively. In conclusion, provided mechanisms are in place to address increased variations of INR at higher ranges, POC testing can be successfully implemented in busy hospital-based anticoagulation clinics.