Anästhesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie : AINS
-
Anasthesiol Intensivmed Notfallmed Schmerzther · May 1992
[Are humidity filters necessary in the inspired air in the breathing circuit? A new in vivo method of measuring humidity in the air breathed].
Humidification of inspiratory gases under anaesthetic conditions still is a matter of controversial discussion. Physiological humidification and heating of breathing air are preconditions for mucociliary clearance, pulmonary cleaning and defence mechanisms. These functions of the upper respiratory tract are eliminated by application of artificial airways. ⋯ After one hour's respiration with this system without HME a relative humidity of 87% at 30 degrees C (26 mg/l) is reached after replaced HME. Initial relative humidity in a semiopen CPAP-system is about 12% at 28 degrees C (3 mg/l). This is increased to 85% at 29.5 degrees C (25 mg/l) after 15 minutes respiration with HME.(ABSTRACT TRUNCATED AT 250 WORDS)