Comparative StudyA comparison between the automatized impedance cardiography and pulsed-wave Doppler echocardiography methods for measurements of stroke volume (SV) and systolic time intervals (STI).
Stroke volume (SV) and systolic time intervals (STI) were measured automatically using impedance cardiography signals (ICG) and compared with those obtained by pulsed-wave Doppler echocardiography using the apex approach. The comparison was made in 9 healthy male subjects, mean age 24.9 +/- 12.2 years, using recordings of 10 heart cycles simultaneously obtained by the two methods. During measurements the subjects rested in the supine position. ⋯ The regression equation for SV was: SVA = 0.784. SVE + 15 (r = 0.69, p < 0.001, SEE = 10.7 ml). We conclude that automatic determination of SV and ET from ICG signals provides results comparable in absolute values with those obtained by the pulsed wave Doppler ultrasonocardiography using the apex approach for subjects remaining in the supine position.