Comparison of three commercially available hollow fiber oxygenators: gas transfer performance and biocompatibility.
The new generation of oxygenators have improved blood flow pathways that enable reduction in priming volume and, thus, hemodilution during cardiopulmonary bypass (CPB). We evaluated three oxygenators and two sizes of venous reservoirs in relation to priming volume, gas transfer, and blood activation. To compare priming volume, gas transfer, and biocompatibility of three hollow fiber oxygenators and two different size venous reservoirs, 60 patients were randomly allocated in groups to undergo cardiopulmonary bypass. ⋯ The oxygenator with the largest blood contact surface area and improved geometric configuration (group 3) showed the lowest oxygen transfer rate per square meter. However, this oxygenator elevated oxygen partial pressure the most and reduced carbon dioxide partial pressure the most. In group 2, where a smaller venous reservoir was used, the highest blood activation was observed.