The HemoCue B-hemoglobin test system (HemoCue, Inc., Mission Viejo, CA) is a photometric method for rapid bedside determination of hemoglobin (Hb). We compared the performance of HemoCue measured Hb against Coulter STK-S (CSTK) measured Hb in chronic hemodialysis (HD) patients in two different settings. In the first setting, Hemocue analysis was performed by multiple HD technicians (n = 132). ⋯ After reviewing the two protocols, the primary difference in the two studies was the technique used to obtain the specimens. When performed properly, Hb testing using the HemoCue testing system had a high level of agreement with CSTK. Appropriate training in specimen handling, as well as test performance, will increase accuracy and reliability of bedside hemoglobinometry.
-
To evaluate the ability of arteriovenous carbon dioxide removal (AVCO2R) to maintain hemodynamic stability during treatment of adult respiratory distress syndrome (ARDS), we used our smoke/burn, LD40 sheep model of ARDS. With onset of ARDS (PaO2/FiO2 < 200) animals were randomized to AVCO2R (n = 20) or SHAM (n = 8). ⋯ At 48 hours, AVCO2R allowed significant reductions (p<0.05) in minute ventilation (13.6+/-2.5 to 7.6+/-0.8 L/min); tidal volume (TV) (389.4+/-24.1 to 295.0+/-10.1 ml); peak inspiratory pressure (PIP) (25.4+/-9.2 to 18.8+/-2.5 cm H2O); RR (27.5+/-0.7 to 21.6+/-1.8 breaths/min); and FiO2 (0.96+/-0.00 to 0.48+/-0.2) while normocapnia was maintained. AVCO2R is an effective method of CO2 removal during severe respiratory failure that is hemodynamically well tolerated.