The effects of pulsatile flow upon renal tissue perfusion during cardiopulmonary bypass: a comparative study of pulsatile and nonpulsatile flow.
This study was conducted to directly compare the effects of pulsatile and nonpulsatile blood flow in the extracorporeal circulation upon renal tissue perfusion by using a tissue perfusion measurement system. A total cardiopulmonary bypass circuit was constructed to accommodate twelve Yorkshire swine, weighing 20 approximately 30 kg. Animals were randomly assigned to group 1 (n = 6, nonpulsatile centrifugal pump) or group 2 (n = 6, pulsatile T-PLS pump). ⋯ The intergroup difference was significant at 30 minutes (47.5 +/- 18.3 ml/min/100 g in group 1 vs. 83.4 +/- 28.5 ml/min/100 g in group 2; p = 0.026). Pulsatile flow achieves higher levels of tissue perfusion of the kidney during short-term extracorporeal circulation. A further study is required to observe the effects of pulsatile flow upon other vital organs and its long-term significance.
-
Ultrasonic flow probes have been used to optimize biventricular pacing immediately after cardiopulmonary bypass, improving cardiac output (CO) by 10%; however, flow probes must be removed with chest closure. The PulseCO system (LiDCO Limited, Cambridge, UK) may extend optimization into the postoperative period, but controlled validations have not been reported. Six anesthetized pigs were instrumented for right heart bypass. ⋯ When mean arterial pressure was increased by 20% using a phenylephrine infusion, PulseCO falsely indicated an increase in CO (2.13 vs. 2.47 L/min, p = 0.014). When mean arterial pressure was decreased by 20% using a nitroprusside infusion, PulseCO falsely indicated a decrease in CO (2.13 vs. 1.79 L/min, p = 0.003). PulseCO appears to be useful for assessing acute changes in CO if its limitations are recognized.
-
Comparative Study
Precise quantification of pressure flow waveforms of a pulsatile ventricular assist device.
Unreliable quantification of flow pulsatility has hampered many efforts to assess the importance of pulsatile perfusion. Generation of pulsatile flow depends upon an energy gradient. It is necessary to quantify pressure flow waveforms in terms of hemodynamic energy levels to make a valid comparison between perfusion modes during chronic support. ⋯ Similar results were obtained when the pump rate was changed from 70 bpm to 80 bpm. The EEP and SHE formulas are adequate to quantify different levels of pulsatility for direct and meaningful comparisons. This particular pulsatile VAD system produces near physiologic hemodynamic energy levels at each experimental stage.
-
In this article, data from the annual international Extracorporeal Life Support Organization (ELSO) Registry Reports through July 2004 are presented. Nearly 29,000 patients have received extracorporeal life support (ECLS). Of the patients, 66% (more than 19,000) were neonatal respiratory failure, with a 77% overall survival to discharge or transfer. ⋯ Survival to discharge, however, is comparable with pediatric respiratory failure (53%). Support continues to increase for cardiac failure across all age groups, with survival to discharge rates of 38%, 43%, and 33% for neonatal, pediatric, and adults, respectively. Survival in pediatric and adult respiratory failure cases supported with ECLS has remained fairly consistent over the past 5 years.